| Model Number |
25058 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure the doctor was suturing in flixene graft when he mentioned that the outer layer of the graft had appeared to have separated.He also commented that the flixene graft he implanted the week prior also had the same issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Additional information also stated: procedure was a thigh implant and graft is still implanted.Surgical delay 5 mins to discuss issue & take photo, no harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Related mfg report number 3011175548-2024-00105 investigation summary: this complaint reports that a the outer layer of a flixene graft (p/n 25058, l/n 471166) separated from the rest while suturing.The user was able to complete the surgery with the graft and there was no harm to the patient other than the procedure being prolonged for a few minutes to take pictures of the layer separation.The user stated that they inspected the graft before use and there was no sign of damage.They confirmed that a 6mm atrium tunneler was used and the sheath was covering the graft while it was being placed.The user cut the graft with sharp surgical scissors.It was while suturing that the separation was first noticed.Two pictures were provided, showing the graft in place and partially sutured with the outer layer being peeled back with surgical tweezers.The procedure was completed without replacing the graft and it remains implanted in the patient.The user stated that this was the second time this had occurred in about a week (the other occurrence is documented in complaint (b)(4).The dhr for lot 471166 and relevant subassemblies were reviewed and no anomalies in manufacturing were found.No ncrs were identified for any of these lots.There is no evidence to suggest that manufacturing, equipment, or design are related to the complaint root cause.The ifu provides adequate instructions for the use of this device and cautions the user on what tools to use while handling and cutting grafts.There is not indication that the ifu is related to this complaint.Complaint trending found that the actual occurrence level did not exceed the anticipated occurrence level.An excursion was identified in april 2024 for flixene grafts exceeding the 3 standard deviation limit for the associated reported defect for this complaint.A complaint history review was completed which found one similar complaints.A recurring lot number report was completed which identified no other complaints involving lot 471166.A review of crs/capas identified no related crs or capas other than cr 1051043, which was opened for the complaint trending excursion in april 2024.An ncr review was completed which found none related to this complaint.This complaint is confirmed based on the pictures the user provided, as is a device nonconformance.However, the information they provided and the information reviewed in this investigation indicates that the most likely root-cause is user error.There is no indication that the product was nonconforming when it left the manufacturing facility or that the instructions for use were inadequate.No damage was seen when the graft was examined prior to implantation and the separation of the outer layer was not identified until after the graft was cut.This likely means that the separation occurred during and due to the cutting, as that is a known cause of layer separation.H3 other text : device not available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data section: h11 after further evaluation of the details provided, it has been determined that the use of a smaller than recommended tunneling rod tip in this procedure likely lead to increased friction while pulling the graft into place which caused the separation that was observed.Per the instructions for use, ¿once the physician has confirmed that the appropriate size graft is available, he can select the appropriate vascular graft tunneler rod and tip.To create the most appropriate size tunnel for the implant, use the following guidelines for tip selection: for ringed vascular grafts or thick walled products such as flixene vascular graft, use the next larger size tip.¿ in this case, the graft was a 6mm diameter so it is recommended for a 7mm tunneler tip to be used.Additionally, the pictures that were provided show the graft being held with a serrated metal clamp.This likely also contributed to the separation of the outer layer.The ifu warnings state ¿clamping of the graft should be avoided whenever possible and limited to clamps shod with soft material.¿.
|
| |
|