Catalog Number 9-ASD-008
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a 8mm amplatzer septal occluder was selected for implant on (b)(6) 2024 using a 7f amplatzer torqvue delivery system to treat the atrial septal defect.A 9mm amplatzer septal occluder was selected for implant using a 7f amplatzer torqvue delivery system.Both devices were deployed into the patient at the same time.During implant the 8mm occluder took on a cobra shape.The device was removed from the patient and replaced with a new 9mm amplatzer septal occluder.During implant the first 9mm occluder took on a cobra shape.The device was removed from the patient and replaced with a new 10mm amplatzer septal occluder.There were no interactions with cardiac structures.There were no angulations or kinks noted in the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.The patient status was stable.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.It was indicated that there was no anatomical interference, there was no angulation or kink in the delivery system upon deployment, and a 7f delivery sheath was used.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.A fluoroscopy cineogram is shown with one aso device already deployed within the inter-atrial septum with traction being applied to the right atrium disc, while a second device is being deployed from another delivery sheath.The second device being deployed shows tendency to form a cobra deformation, while the first device already deployed within the inter-atrial septum shows a normal configuration with the right atrial disc bulging due to traction.The cause of the reported incident could not be conclusively determined.However, per the instructions for use, the recommended the sheath size delivery system for use with a 8 mm amplatzer septal occluder is an 6f, and use of a larger than recommended sheath could have contributed to the device deformation.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19088633
Report Number2135147-2024-01645
Device Sequence Number7819394
Product Code MLV
UDI-Device Identifier00811806010052
UDI-Public(01)00811806010052(17)271031(10)8775813
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9-ASD-008
Device Lot Number8775813
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/18/2024
Supplement Date Received by Manufacturer04/23/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
7F AMPL TORQ DS [9-ITV07F45/80, LOT: 8984706]
Patient SexUnknown
-
-