| Catalog Number |
INS096009401 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a reverse total shoulder replacement surgery, the knob of the rss humeral body inserter/extractor knob broke off.It is unknown how the procedure was completed or if there was any delay.No further consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.However, the photograph was reviewed and revealed that the tip of the device is fractured.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device finds the device returned heavily worn from use, with the shaft fractured away just below the hub, the shaft was not returned.The laser markings are degraded.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: additional information ¿d4: lot number¿.
|
| |
|