ARTHREX, INC. SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on 8/24/2021 by a distributor rep that on (b)(6) 2021 two peek knotless suture-tak suture anchors were being used in a case.During use, both anchors failed and did not work.No further information, additional details requested.Additional information provided 8/27/2021: procedure being performed was an ankle instability repair.The tunnel was not made correctly prior to the insertion of the anchors so when pulling the sutures, the placement of the anchors was loose.The tunnels were prepared with a shaverdrill for 3.0 mm knotless suturetak, hard bone ar 1938dsr.The procedure was completed with an ar-1915snf.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19088838
Report Number1220246-2024-01980
Device Sequence Number7112686
Product Code MAI
UDI-Device Identifier00888867024229
UDI-Public00888867024229
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K180594
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberSUTR ANCH PEEK S-TAK KNOTLESS 3X12.7MM
Device Catalogue NumberAR-1938PS
Device Lot Number12162969
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2021
Initial Report FDA Received Date04/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-