• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LINR, NTRL, 36MM ID, GROUP 4 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LINR, NTRL, 36MM ID, GROUP 4 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 130-36-54
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Osteolysis (2377); Insufficient Information (4580)
Date of Event 11/02/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10.D10.Concomitants: 4148874, 164-01-14 - element-stem, collarless w/ha, std offset, sz 14.4653649, 170-36-00 - biolox delta femoral head 36mm od, +0mm.4617767, 170-36-03 - biolox delta femoral head 36mm od, +3.5mm.4488386, 186-01-60 - integrip cc, cluster 60mm, g4.Presumed left hip -4415175, 130-36-54 - nv gxl linr, ntrl, 36mm id, group 4 cups, serial number (b)(6) is confirmed to have been packaged in a vacuum bag that does not contain evoh.H3.Investigation results: the nv gxl linr, ntrl, 36mm id, group 4 cups device with serial number (b)(6) is confirmed to have been packaged in a vacuum bag that does not contain evoh.The cause of the patient's condition and the reason for the planned revision are not reported.There is insufficient information to determine the relationship of the devices to the patient's condition at this time.When more information is received it will be evaluated and additional actions taken as appropriate.These devices are used for treatment not diagnosis.H6.Health effect - impact code is for pending surgical revision.There is no other information available.
 
Event or Problem Description
It was reported via legal documentation that a patient had a total right hip arthroplasty on (b)(6) 2016, with a planned revision, as of (b)(6) 2024 the revision date and surgical information have not been received.Scheduled date for revision was reported to be (b)(6) 2023.There is no other information provided/available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: d7a.The following sections were corrected: b1 b2 g2 h6.The most likely cause for the reported revision due to prosthesis wear and osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINR, NTRL, 36MM ID, GROUP 4 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
matt collins
MDR Report Key19089286
Report Number1038671-2024-00808
Device Sequence Number7071573
Product Code LZO
UDI-Device Identifier10885862022240
UDI-Public10885862022240
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2021
Device Catalogue Number130-36-54
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2023
Supplement Date Received by Manufacturer02/07/2025
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-