BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00517470
Medical Device Problem Codes Break (1069); Activation, Positioning or Separation Problem (2906); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.Block h6: imdrf device code a15 captures the reportable event of stent partially deployed.
 
Event or Problem Description
It was reported to boston scientific corporation that an agile esophageal tts fuly covered stent was to be implanted in the esophagus to treat a malignant stricture during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2024.During the procedure, the sheath broke during attempted deployment.The stent was removed from the patient partially deployed and another agile esophageal stent was used to complete the procedure.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.Block h6: imdrf device code a15 captures the reportable event of stent partially deployed.Block h10: an agile esophageal stent and delivery system were received for analysis.Visual examination of the returned device found the stent partially deployed, the outer sheath detached and the inner sheath and stainless steel kinked.No other problems were noted to the stent and delivery system.The reported events of stent partially deployed and sheath break can be confirmed.Taking all available information into consideration, the investigation concluded that the reported events and the observed damages were likely due to factors encountered during the procedure.It may be that lesion characteristics, handling of the device and the technique used by the physician (force applied), limited the performance of the device and contributed to the reported events and observed damages.Therefore, a review and analysis of the all the available information indicated that the most probable cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that an agile esophageal tts fuly covered stent was to be implanted in the esophagus to treat a malignant stricture during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2024.During the procedure, the sheath broke during attempted deployment.The stent was removed from the patient partially deployed and another agile esophageal stent was used to complete the procedure.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILE ESOPHAGEAL
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key19089460
Report Number3005099803-2024-01682
Device Sequence Number7576262
Product Code ESW
UDI-Device Identifier08714729973096
UDI-Public08714729973096
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K180144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00517470
Device Catalogue Number1747
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2024
Supplement Date Received by Manufacturer04/30/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date05/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-