• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC CXI SUPPORT CATHETER; KRA CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC CXI SUPPORT CATHETER; KRA CATHETER, CONTINUOUS FLUSH Back to Search Results
Catalog Number CXI-2.6-18-90-ANG
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: customer name and address = postal code: (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, during an unknown procedure, an unspecified 0.018-inch wire perforated a cxi support catheter.The catheter was advanced through an unknown 4-french sheath.An ipsilateral approach was used, and the anatomy was very tortuous.At the beginning of the procedure, before crossing the stenosed area, the wire perforated the catheter on advancement, exiting the top ("anterior tip") of the catheter rather than the exit hole at the catheter tip.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during an unknown procedure, an unspecified 0.018-inch wire perforated a cxi support catheter.The catheter was advanced through an unknown 4-french sheath.An ipsilateral approach was used, and the anatomy was very tortuous.At the beginning of the procedure, before crossing the stenosed area, the wire perforated the catheter on advancement, exiting the top ("anterior tip") of the catheter rather than the exit hole at the catheter tip.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.A puncture of the catheter material was found approximately 6 centimeters from the distal end of the strain relief.A document-based investigation evaluation was performed.A device history record review of the complaint lot found no relevant non-conformances.A review of the device history record for a component lot associated with the complaint lot found one relevant non-conformance on two devices; however, all affected product was scrapped prior to release.A review of complaint history found no additional complaints for this lot number.The customer followed all relevant instructions in the ifu related to this failure.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.Although one relevant non-conformance was noted on a component lot, all non-conforming product was scrapped, there are 100% inspections in place to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Therefore, it was concluded that there is no evidence that additional non-conforming product exists in house or in the field.Based on the information provided and the results of the investigation, cook has concluded that the patient's anatomy contributed to the device failure in this incident, as the complaint device was punctured by the wire guide in "very tortuous" anatomy.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CXI SUPPORT CATHETER
Common Device Name
KRA CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key19089640
Report Number1820334-2024-00508
Device Sequence Number18055068
Product Code KRA
UDI-Device Identifier00827002565517
UDI-Public(01)00827002565517(17)260901(10)15638114
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K122796
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2026
Device Catalogue NumberCXI-2.6-18-90-ANG
Device Lot Number15638114
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2024
Supplement Date Received by Manufacturer06/27/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-