| Catalog Number |
9-PFO-3525 |
| Medical Device Problem Codes |
Difficult to Fold, Unfold or Collapse (1254); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on 22 march 2024, a 35-25mm amplatzer talisman pfo occluder was selected for an implant.During the procedure, the device could not be deployed from the cable.It is unknown if a device was implanted.The patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
¿ investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of separation problem could not be confirmed.The returned device was attached and detached from the the delivery cable without any issues.Additionally, the returned device was observed to be deformed when it was received.However, the returned device went back to its proper conformation after decontamination.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the available information, the cause of the reported separation problem could not be conclusively determined.The observed material deformation could not be determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|