| Model Number |
16171111 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2024 a transforaminal lateral interbody fusion was conducted at l4/s1.On (b)(6) 2024 it was discovered the screw shank separated at l4.No patient injuries were reported.No revision reportedly performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is in-situ, but images provided confirmed the complaint.Review of the images provided identified the tulip head on the left cranial side screw detached from the shank head.No root cause could be determined however review of the reported event, associated technique guide, complaint review, and radiographs provided suggest an incomplete assembly in-situ which can be the result of technique or anatomical obstruction.No additional investigation can be completed at this time.Should the device be returned for evaluation another record will be filed.Labeling review: "potential adverse events and complications as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation." "warnings, cautions and precautions.The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants." "patient education preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "pre-operative warnings.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "post-operative warnings.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." 9401879-en p-12/2018 h3 other text : device in-situ.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received by nuvasive and the complaint was confirmed.Examination revealed the tulip was not attached to the shank with minor scratches and scrapes on the side of the ball head as well as a flat dent (chamfer) at the edge of the internal hex from forced contact.The tulip was inspected where a deep gouge was presented on the interior tulip surface adjacent to the shank port as well one of the split ring tips is fractured off and jammed caudally with an increase split gap over double the required amount.The damage observed indicates the screw was misaligned during the attempted in-situ assembly allowing the shank head to contact and damage both the tulip port surface and split ring causing damage to the retention features and subsequent separation and considered the root cause of the implant damage requiring revision.No additional investigation required.Manufacturing review: review of the device history records of all devices notes no material non-conformance¿s, no manufacturing errors, nor discrepancies with respect to material type, treatments, dimensions that may have caused or contributed to this mode of failure.Parts met acceptance criteria upon release.Labeling review: "potential adverse events and complications as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation." "warnings, cautions and precautions ¿the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants." "patient education preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "pre-operative warning¿care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "post-operative warnings¿damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." 9401879-en p-12/2018.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024 a patient underwent a transforaminal lumbar interbody fusion procedure from l4 to s1 and the surgery was completed without issue.On (b)(6) 2024 it was reported that it was discovered that the left l4 tulip had separated from the shank.No adverse patient consequences have been reported and no revision has been reported at this time.On (b)(6) 2024 a revision took place where the shank and tulip were removed.New or updated information is listed in h10.
|
| |
|