Catalog Number GTPOSTF1
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Investigation results: returned 1x gtpostf1 drill.  checked under microscope and found no manufacturing marks or defects.  see attached photo, break occurred between the handle and the 1st calibration line.  unit setting and number of uses is unknown.  no other complaints for this lot number.  no unopened product was returned therefore no further testing can be performed and root cause is unknown.Dhr: nothing unusual to report was found during dhr review.A query indicates no additional complaints have been received for this lot number.
 
Event or Problem Description
In this event it is reported that gt fiber post 1mm - post drill broke during use.The broken pieces were retrieved and no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GT FIBER POST 1MM
Common Device Name
POST, ROOT CANAL
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key19091500
Report Number2320721-2024-00021
Device Sequence Number7071640
Product Code ELR
UDI-Device IdentifierD716GTPOSTF11
UDI-PublicD716GTPOSTF11
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberGTPOSTF1
Device Lot Number0000392513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2024
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/02/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/02/2024
Initial Report FDA Received Date04/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-