SYNTHES GMBH DHS/DCS COUPLING SCREW; EXTRACTOR Back to Search Results
Catalog Number 338.31
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2024, during the procedure, the surgeon was attempting to remove a dhs one-step lag screw with the dhs/dcs one step insertion wrench and dhs/dcs coupling screw (for one-step insertion wrench).After applying increased force in attempting to rotate the wrench counterclockwise, the surgeon suddenly felt no resistance and the wrench, when turned counterclockwise, resulted in no progress of the lag screw being removed.Upon inspecting the coupling screw, which was properly threaded into the lag screw initially, it was determined that the threaded portion of the coupling screw had broken off and remained threaded in the lag screw.In addition, it was determined that the upper portion of the insertion wrench (the "t") had lost its connection to the shaft.If this is a weld between the "t" portion of the wrench and the shaft, it was that weld that had become disconnected.This resulted in the upper "t" portion rotating when turned without the shaft of the wrench turning.All hardware was eventually removed and the procedure was successfully completed.There was no patient consequences outcome.This report is for one (1) dhs/dcs coupling screw this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D2: date of concomitant therapy is (b)(6) 2024.H6: part: 338.31.Synthes lot: h731988.Supplier lot: h731988.Release to warehouse date: 25 august 2019.Supplier: avalign technologies nemcomed.No nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H6 component code: g07002 ¿ device not returned.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DHS/DCS COUPLING SCREW
Common Device Name
EXTRACTOR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19092644
Report Number8030965-2024-04801
Device Sequence Number17917905
Product Code HWB
UDI-Device Identifier10886982193322
UDI-Public(01)10886982193322
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number338.31
Device Lot NumberH731988
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer04/13/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DHS/DCS ONE-STEP INSERTION WRENCH; UNK - SCREWS: DHS/DCS
Patient SexUnknown
-
-