• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 S/+0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 S/+0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 75100856
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a thr procedure, the inner tabs of a trial femoral head 36 s/+0 were broken and did not hold onto stem as designed.No pieces fell into patient.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during a total hip replacement procedure, the inner tabs of a trial femoral head 36 s/+0 were broken and did not hold onto stem as designed.Patient was not injured as consequence of this problem.The complaint device was returned for investigation.A visual evaluation of the device was conducted, and it was concluded that there is a fracture at one of the inner flaps.Furthermore, there are signs of wear such as scratches and dents.Apart from that no further defects were detected.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The review of historical complaints for the alleged device revealed no additional similar complaints reported for the same batch, and 14 additional similar complaints for the same product number over the past 12 months with similar failure mode.The root cause of the event can be attributed to a known inherent risk of the device.Previous investigations demonstrated that the trial femoral head may disconnect from the stem taper intra-operatively and flaps might bend or break off.These failure modes may further be exacerbated by excessive use of the device over time.The performance of the device is still within the risks level that are anticipated in the risk management documentation.The current design will be further monitored through post-market surveillance processes.No further actions are necessary at this time point.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.The root cause of the event can be attributed to a known inherent risk of device.Considering the performed visual inspection, it is suspected that the mentioned device failure arose from a ¿wear and tear" issue.Normal wear and tear are known to contribute to the reported event.Additionally, repeated steam sterilization processes could contribute to an embrittlement.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.The need for further actions is not indicated.Nevertheless, smith+nephew will continue to monitor this device for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL FEMORAL HEAD 36 S/+0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau TN CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau TN CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19093001
Report Number9613369-2024-00037
Device Sequence Number15229536
Product Code LZO
UDI-Device Identifier07611996121896
UDI-Public07611996121896
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K132435
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75100856
Device Lot NumberA73038
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2024
Supplement Date Received by Manufacturer04/24/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date04/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-