| Model Number |
3100A VENT,115V,60HZ,DOM |
| Medical Device Problem Code |
No Pressure (2994)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that while attempting to adjust setting to map of 9cmh2o and a delta p of 24 on the 3100a the vent stops.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.However, customer requested a quote to get field service out to evaluate the problem.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : on site- repair.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of on-site repair: lowered set max to 40 and lower and was to lower map below 9.
|
| |
|