The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the lcb (light carrying bundle) fluid damage, the operation channel (primary) perforated, the control body corroded, the lg connector corroded, the operation channel (primary) leak, and the distal body worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|