|
Failure analysis investigations confirmed the customer-reported complaint.The biopsy needle was found to have a needle retraction failure.It was manually extended but could not be retracted.During in-house testing, the needle stop was pressed down and slid down the ratchet indicator to position #3.The grey handle was moved distally to extend the needle out of the sheath tip by 3cm.However, the needle could not be manually pushed back into the sheath or pulled back into it.The sheath movement was also not able to be performed due to the biopsy needle retraction failure.Visual inspection revealed no physical damage or kinks on the needle and sheath.The sheath length was measured to be 926mm, which falls within the 925mm +/- 1mm sheath length specification.Failure analysis identified a secondary finding of a damaged stylet tip unrelated to the customer-reported complaint.The biopsy needle was also found with a small metal fragment, approximately 2mm in length, which is likely from the damaged stylet tip.
|
|
The biopsy needle 1.5 has been returned and evaluated by the advanced failure analysis team on 05/24/2024.An approximately 1.8mm by 0.05-0.26mm fragment was returned with the device.The size of the fragment was observed to be larger than the area of the nitinol stylet that was referred to as a "damaged stylet tip" during primary failure analysis; fragment does not appear to have been generated from the stylet.The area of the stylet tip that appears to slightly worn, does not appear to have any sharp edges or obvious missing pieces.The distal tip is deburred at the supplier, and no burrs, sharp edges, or loose flash is permitted per the print.Rma was reviewed with ion biopsy needle manufacturing engineer, ion biopsy needle design engineer, and ion biopsy needle mechanical engineer of manufacturing equipment.Pin gauge was used to measure the id of the sheath tip.Sheath tip id was measured to be approximately 0.038in, which is within the 0.0380in +/- 0.0005in specification.Needle assembly was disassembled to check components for damage or missing fragments.No damage was observed to the needle tip, sheath tip, or flexible collar.Fragment does not appear to have come from the needle assembly.Biodebris was observed along the needle and needle shaft.Per ion biopsy needle manufacturing engineer, the only potential impact from the manufacturing line leading to a failure at this location could be if the proximal portion of the needle shaft got kinked by the glass mold, and the kink was not detected by the technician.Upon measuring the length of the end of the glass mold from the taper of the shaft bond, a potential manufacturing induced kink would be approximately 0.98in from the distal end of the taper.The shaft break of the rma un occurred approximately 14.7mm (or 0.58in) from the taper, which is approximately in the middle of the portion of the needle inside the glass mold.Therefore, it does not appear that a kink from the glass mold during manufacturing is the cause of the failure.Per ion biopsy needle design engineer, bond failure was unable to be recreated in a simulated use high friction scenario in a koken model.Further evaluation was performed in an attempt to replicate the shaft failure.Upon securing the an in-house needle at the needle shaft to flexible needle bond and pulling until failure of the shaft, significance necking was observed.A second in-house needle shaft was kinked at the location of the failure observed in the returned unit to simulate a worst-case manufacturing scenario of the needle shaft becoming kinked during manufacturing.Upon securing the needle at the needle shaft to flexible needle bond and pulling until failure of the shaft, significant necking was observed despite the kinked shaft.Bond failure with no necking of the shaft similar to what was observed on the rma unit was not able to be recreated.No necking was observed on the rma unit need shaft, and the point of failure is in a normally unstressed region.It is possible that the aluminum fragment was embedded inside the needle shaft at or near the point of failure, and that flexing within the catheter and friction with the stylet caused the fragment to cut into the needle shaft, weakening the shaft in this location.If the needle shaft were weakened then the force required to cause the failure of the needle shaft would be low, which may explain the lack of necking observed on the shaft.Root cause of the shaft breakage is most likely attributed to a component susceptibility to damage of the needle shaft.Fragment was sent to an external lab for analysis to determine chemical composition.External lab analysis determined that the fragment is aluminum.Needle shaft was observed to be broken at the proximal end of the heat affect zone of the needle tip bond.No necking of the shaft was observed.Per mechanical engineer of manufacturing equipment, there are several efts in the clean room that are comprised of custom machined aluminum parts.The fragment appears similar to a metal shaving from a machining operation of an eft; however, shavings are cleaned from the part during the finish or cleaning process.Furthermore, the efts are wiped down prior to entry into the clean room and on a regular basis.
|