| Model Number |
SUTURE ANCHOR, BIO-CORKSCREW FT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopy surgery of the shoulder joint with the remplissage method the implant broke while being screwed into a pre-drilled hole.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is not confirmed.Visual inspection identified that the implant 5.5mm of the suture anchor, biocomposite corkscrew® ft,vented 5.5 x 14 mm with #2 fib erwire® and #2 tigerwire® had not returned or picture provided to verify the allegation.The sutures were not returned.Functional testing was not performed due to the damage to the device.The method of bone preparation and the quality of the bone encountered were not provided.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|
Search Alerts/Recalls
|