U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The diamondback 360 coronary orbital atherectomy device (oad) was used to treat a 2.5mm, 90% stenosed, heavily calcified with mild tortuosity left anterior descending artery (lad).The oad was spun for one treatment on low speed and a perforation was observed in the mid lad.The patient declined, cardiopulmonary resuscitation (cpr) was performed for twenty to thirty minutes while simultaneously performing pericardiocentesis and extracorporeal membrane oxygenation (ecmo).One liter of blood was drained from the heart.Three non-csi/abbott balloons and three non-csi/abbott stents were used and placed to successfully treat the perforation.Flow was restored and the patient was stable and placed on a ventilator and transferred to the intensive care unit (icu).The following day the patient's dependance on the ventilator was reduced.It was the physician's opinion the oad caused the perforation.
 
Additional Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for user manual states that a perforation of vessels is a potential adverse event that may occur and/or require intervention with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Common Device Name
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
poonoy chanthavongsa
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key19095968
Report Number3004742232-2024-00157
Device Sequence Number18068737
Product Code MCX
UDI-Device Identifier10850026568087
UDI-Public(01)10850026568087(17)250831(10)504173-1
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P130005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberDBEC-125
Device Catalogue Number7-10060-09
Device Lot Number504173-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Initial Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight62 KG
-
-