| Medical Device Problem Codes |
Mechanical Problem (1384); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
03/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The gi4000 electro surgery unit subject of the reported event is being returned to the manufacturer for evaluation.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure, user facility personnel were utilizing the argon plasma coagulation on the gi4000 electro surgery unit and the display screen that shows the endoscope image "blacked out".The patient procedure was subsequently cancelled.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The gi4000 electro surgery unit subject of the reported event was returned to the manufacturer for evaluation.Evaluation of the device determined that no issues were noted with the function or operation.The device operated according to specifications and the reported event could not be duplicated.During post-test inspection, it was noted that a "noise" was coming out of the argon regulator.The manufacturer notified the user facility of this observation; however, the user facility declined further repair and requested the gi4000 electro surgery unit be returned to them.The unit has been returned and put back in service.The device history record (dhr) was reviewed and confirmed the device was manufactured to specification; no abnormalities were noted.A complaint review was performed and confirmed this event to be isolated.No additional issues have been reported.
|
| |
|
Search Alerts/Recalls
|