• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 58MM G; PROTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 58MM G; PROTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 04/03/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A2: date of birth: information provided by the hospital showed the date of birth documented as two different dates: (b)(6) 1947 and (b)(6) 1948.D10: cat# 010000985 lot# 7648566 g7 freedom const e1 lnr 36mm g.Cat# 11-107021 lot# 884560 freedom constr hd 36mm t1 +9mm.Cat# 010000997 lot# 7681145 g7 screw 6.5mm x 20mm.Cat# 010000997 lot# 7641778 g7 screw 6.5mm x 20mm.Cat# 010000999 lot#7665793 g7 screw 6.5mm x 30mm.Cat# 010000999 lot# 7633835 g7 screw 6.5mm x 30mm.Cat# 010000999 lot# 7648566 g7 screw 6.5mm x 30mm.G2: foreign: country: australia.Product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient had a right total hip arthroplasty on an unknown date.The patient had a right hip revision approximately one month ago due to recurrent dislocations.Subsequently, the patient was revised approximately a weeks ago due to a pelvic fracture which led to cup loosening.It was reported that the pelvic fracture was thought to have occurred in the previous revision.No additional information available.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;g3;h2;h3;h6 no product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The reported products, shell, liner, and screws were reviewed for compatibility with no issues noted.Medical records were not provided for the event related to the revision due to pelvic fracture leading to cup loosening.However, review of the available information by a health care professional identified the following: it was reported that patient had a right hip performed on an unknown date.Patient subsequently had a right hip revision due to recurrent dislocations and eccentric poly wear and was then revised a second time for recurrent dislocations.The patient experienced a pelvic fracture leading to cup loosening which resulted in an additional revision.Contributing factors to the bone fracture is unknown.The shell, liner and screws were explanted, and an unknown zimmer biomet stem was retained.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 58MM G
Common Device Name
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19097091
Report Number0001825034-2024-01005
Device Sequence Number7428827
Product Code LPH
UDI-Device Identifier00887868356056
UDI-Public(01)00887868356056(17)340111(10)66390801
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110010267
Device Lot Number66390801
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2024
Supplement Date Received by Manufacturer07/30/2024
Initial Report FDA Received Date04/12/2024
Supplement Report FDA Received Date07/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-