| Catalog Number |
234-020-023 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an incident where the product has snapped during an operation.The broken pieces were potentially unable to be retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device broke.Probable root cause: process.Instrument manufactured out of specification.Instrument not assembled properly.Application.User not applying forces correctly on instrument.User applying excessive force on instrument.Use past expected device lifetime.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an incident where the product has snapped during an operation.The broken pieces were potentially unable to be retrieved.
|
| |
|
Search Alerts/Recalls
|