| Model Number |
MAF-GM2 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the airway mobilescope exhibited peeling off the coating on the insertion section.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: h8.Additional information added to field h3,h4,h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation and the information provided, the cause of the event was unable to be specified.Presumably, the defect was caused by stress of repeated use, external factors, or handling of the device.The following is included in the instructions for use (ifu): instructions, operation manual chapter 3 preparation and operation¿ section 3.6 ¿inspection of the endoscope¿ describes how to inspect for the subject event as below.Olympus will continue to monitor field performance for this device.
|
| |
|