• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY MEGATRON; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY MEGATRON; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0032117237
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the balloon became stuck in the stent struts.The target lesion was located in a non-tortuous and calcified left main (lm) artery.A 5.00 x 16mm synergy megatron drug-eluting stent was advanced to treat the lesion.Following stent deployment, the balloon was deflated and pulled back part way to post-dilate the proximal edge in the lm and use the proximal balloon as a proximal optimization technique (pot) technique.The balloon was deflated normally but got caught on the stent struts as it was deflated and became stuck.A 30mm balloon was inserted in the guide alongside the stent shaft and pinned the stent shaft against the inner guide.The system was then pulled out successfully.The stent was not affected and post-dilatation was performed with a new balloon to complete the procedure.The balloon was removed in a fully deflated state.The procedure was completed and there were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY MEGATRON
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19098579
Report Number2124215-2024-20262
Device Sequence Number15407100
Product Code NIQ
UDI-Device Identifier08714729985648
UDI-Public08714729985648
Combination Product (Y/N)Y
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/10/2025
Device Lot Number0032117237
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/18/2024
Initial Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-