|
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H10: additional manufacturer narrative: updated sections: d4 (expiration date), g3, g6, h2, h4, h6 (type of investigation, investigation findings, and investigation conclusions).The device history record (dhr) review was completed and there were two non-conformances found related to this serial number.However, the valve was re-worked to completion, passing all inspections.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
|