• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LLC MAHURKAR; CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LLC MAHURKAR; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number 8888135162
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during saline pre-fill, there was a leak at the extension tube (arterial side) and bifurcation connection.Nothing unusual was observed on the device prior to use.Flushing/priming was done prior to use and no problem was found.Tego was not utilized and there was no luer adapter problem.Iodophor was used as the cleaning agent on the device.Alcohol-free sanitized iodophor was not utilized to clean the adapters and no ointment was applied.The insertion site treated with prior to product placement.Catheter was not repaired.There were no other products being utilized with the device.No excessive force applied on the device and no instrument was used to loosen or tighten the device.The catheter was explanted and replaced with a product from another manufacture as a remedial action to resolve the issue.There was a 50 milliliter (ml) blood loss and blood transfusion was not required.There was no intervention/treatment provided to the patient.There was no reported patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that there was a cut/hole/disconnection in the extension tube which led to a leak.It was reported that there was a leak or hole on the extension tube at or near the bifurcate/hub.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19099663
Report Number1282497-2024-00039
Device Sequence Number7267935
Product Code MPB
UDI-Device Identifier10884521006379
UDI-Public10884521006379
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K192302
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888135162
Device Catalogue Number8888135162
Device Lot Number2234800115
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2024
Supplement Date Received by Manufacturer10/08/2024
Initial Report FDA Received Date04/12/2024
Supplement Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1