|
|
| Catalog Number |
SASMH-1-S |
| Medical Device Problem Code |
Failure to Conduct (1114)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
510(k): k173673 investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.One possible cause of this report is use with an incompatible active cord.The ifu states "this device must only be used with an active cord compatible with a 3 mm diameter connector.This device has only been verified to be compatible with the following cook active cords: acu-1 and acu-1-vl (supplied non-sterile)." the instructions for use contain the following additional information to assist with proper set-up and use of the device: "before using this device, follow the recommendations provided by the electrosurgical unit manufacturer to ensure patient safety through the proper placement and utilization of the patient return electrode.Ensure that a proper path from the patient return electrode to the electrosurgical unit is maintained throughout the procedure." "inspect the active cord.The cord must be free of kinks, bends, breaks and exposed wires to allow for the accurate transfer of current.If an abnormality is noted, do not use the active cord." "securely connect the active cord to the device handle and electrosurgical unit.The active cord fittings should fit snugly into both the device handle and electrosurgical unit.Following the instructions from the electrosurgical unit manufacturer, position the patient return electrode and connect it to the electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify the desired settings and activate the electrosurgical unit.Note: the maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a colonoscopy, the physician used a cook acusnare polypectomy snare soft.It was reported that the diathermy did not work.There was also no power transmitted for proposed polypectomy.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: the product said to be involved was returned in a white plastic bag with an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved could not confirm the report as it was described.The device was returned with the snare retracted into the sheath.The device would advance and retract as expected when the handle was manipulated, with some resistance.There was a bend in the sheath at the base of the handle (reference medwatch report 1037905-2024-00215).There was also a kink in the catheter noted at 106.8 cm from the base of the handle.The continuity from the electrical pins to the snare head was tested with an ohm meter and passed.An additional functional test with the device was performed by attaching the active cord to the electrical pin.The active cord connected to the device easily and remained securely connected.The device was connected to a valley lab generator and power was applied.The snare cut and coagulated simulated tissue as expected.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the device could not confirm the complaint as it was described, there were no electrical issues present with the snare during our evaluation.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.One possible cause of this report is use with an incompatible active cord.The ifu states "this device must only be used with an active cord compatible with a 3 mm diameter connector.This device has only been verified to be compatible with the following cook active cords: acu-1 and acu-1-vl (supplied non-sterile)." the instructions for use contain the following additional information to assist with proper set-up and use of the device: "before using this device, follow the recommendations provided by the electrosurgical unit manufacturer to ensure patient safety through the proper placement and utilization of the patient return electrode.Ensure that a proper path from the patient return electrode to the electrosurgical unit is maintained throughout the procedure." "inspect the active cord.The cord must be free of kinks, bends, breaks and exposed wires to allow for the accurate transfer of current.If an abnormality is noted, do not use the active cord." "securely connect the active cord to the device handle and electrosurgical unit.The active cord fittings should fit snugly into both the device handle and electrosurgical unit.Following the instructions from the electrosurgical unit manufacturer, position the patient return electrode and connect it to the electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify the desired settings and activate the electrosurgical unit.Note: the maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Search Alerts/Recalls
|
|
|