• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE; FDI, SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE; FDI, SNARE, FLEXIBLE Back to Search Results
Catalog Number AS-1-S
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
510(k): k173673.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.One possible cause of this report is use with an incompatible active cord.The ifu states "this device must only be used with an active cord compatible with a 3 mm diameter connector.This device has only been verified to be compatible with the following cook active cords: acu-1 and acu-1-vl (supplied non-sterile)." the instructions for use contain the following additional information to assist with proper set-up and use of the device: "before using this device, follow the recommendations provided by the electrosurgical unit manufacturer to ensure patient safety through the proper placement and utilization of the patient return electrode.Ensure that a proper path from the patient return electrode to the electrosurgical unit is maintained throughout the procedure." "inspect the active cord.The cord must be free of kinks, bends, breaks and exposed wires to allow for the accurate transfer of current.If an abnormality is noted, do not use the active cord." "securely connect the active cord to the device handle and electrosurgical unit.The active cord fittings should fit snugly into both the device handle and electrosurgical unit.Following the instructions from the electrosurgical unit manufacturer, position the patient return electrode and connect it to the electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify the desired settings and activate the electrosurgical unit.Note: the maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a polypectomy in the colon, the physician used a cook acusnare polypectomy snare.It was initially reported that since the change to the handle of the snare, the snare no longer works smoothly.Outside the patient the snare feels the same as usual, but in the scope/patient the snare works stiffly.Because the snare is not working smooth, the nurse has less control and can¿t work less accurately.Previously, nurses could work with the snare with their eyes closed, but since the change they need 2 hands to handle the snare.This was initially considered to be user preference and was not reportable.Additional information was received on 20 mar 2024 stating that it looks like the cutting is not working properly which is reportable.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSNARE POLYPECTOMY SNARE
Common Device Name
FDI, SNARE, FLEXIBLE
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key19101310
Report Number1037905-2024-00218
Device Sequence Number7343579
Product Code FDI
UDI-Device Identifier00827002226296
UDI-Public(01)00827002226296(17)260712(10)W4745034
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/12/2026
Device Catalogue NumberAS-1-S
Device Lot NumberW4745034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/20/2024
Initial Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCOPE, UNKNOWN MAKE AND MODEL
Patient SexUnknown
-
-