| Model Number |
20226 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent dislodgement occurred.A 9.0x30x75cm express ld iliac / biliary stent balloon was advanced for treatment.However, during the preparation, it was found that the stent was not attached to the delivery system when pulling out from the loop.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that stent dislodgement occurred.A 9.0x30x75cm express ld iliac / biliary stent balloon was advanced for treatment.However, during the preparation, it was found that the stent was not attached to the delivery system when pulling out from the loop.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.Only the stent of the device was returned, and the balloon catheter was not returned.A visual examination found the stent to be damaged/misshapen.No other issues were noted with the stent.
|
| |
|