| Catalog Number |
B33276 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The complaint/event occurred on an unspecified date and involved a 6" smallbore trifuse ext set w/3 microclave®, 0.2 micron filter, 3 clamps, rotating luer.The reporter stated that total parenteral nutrition (tpn) leaked from the device filter during use/infusion from a peripheral intravenous access point on an infant.The trifuse was infusing tpn, lipids and medication.The patient's linen was saturated and the trifuse was found as 'sticky' to the touch.There was no report of human harm associated with the complaint/event.
|
| |
|
Additional Manufacturer Narrative
|
|
One used b33276 smallbore trifuse extension set was returned for inspection/complaint investigation.As received, the filter vents on the 0.2 micron filter were wetted out with prior infusate.The extension set was leak tested per product specifications.There was leakage from the filter vents of the 0.2 micron filter.The reported complaint can be confirmed.The probable cause of leakage is due to a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional/updated information can be found in b3, b5 and e4.Section e additional contact information: (b)(6).Risk manager.
|
| |
|
Event or Problem Description
|
|
A medwatch mandatory report with uf/importer report # (b)(4) was received on 02-may-2024.
|
| |
|
Additional Manufacturer Narrative
|
|
The corrected information can be found in h10.
|
| |
|