| Model Number |
VTICMO 13.2 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm vticmo_13.2 implantable collamer lens of -16.0/2.5/072 (sphere/cylinder/axis) was implanted into the patients left eye (os) on (b)(6) 2024.Lens rotation was observed.On (b)(6) 2024 the lens was explanted and the problem was resolved.Cause reported as patient related factor.The patient has abnormal anatomical structure of the eyes.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d4- expiration date corrected to (b)(6) 2026 claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: h3-lens returned in a microcentrifuge vial; dry with residue/debris on lens.Visual inspection found haptic bent.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|