MAUDE Adverse Event Report: ENCORE MEDICAL L.P ENCORE REVERSE SHOULDER HUMERAL SOCKET REAMER, SMALL PF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P ENCORE REVERSE SHOULDER HUMERAL SOCKET REAMER, SMALL PF Back to Search Results
Catalog Number 804-06-005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Limb Fracture (4518)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to fracture.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
He agent reported that (humerus was fractured due to small socket reamer being too large for this female patient.Had to go to size 6mm x 108mm stem and cement it due to fracture from small shell reamer.Female 65).This event occurred during surgery, near the patient.Significant adverse event was reported by the surgeon.The surgery was completed as intended, with a forty-five-minute delay.The instrument was inspected prior to use and was deemed acceptable for use based on its appearance.The agent was present during surgery and was able to source a suitable replacement device.Broken/fragmented pieces were left in the patient.Rma examination: the reported instrument was not returned to djo surgical for evaluation.The small humeral socket reamer was left in patient per product feedback form.No further evaluation can be made for this event.This customer complaint will be closed.If the device is returned later, the complaint will be updated.The revision level or lot number were not reported; therefore, this instrument could not be linked to a specific device history record (dhr) or the actual date of manufacture could not be determined with confidence.Complaint database review showed thirteen previous complaints but there were no indications that this instrument has a design or material deficiency.The issue in this complaint is not of the instrument but the incorrect size of the reamer used on the patient causing damage to the humerus.This is not an event associated with a product failure, malfunction, or issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE REVERSE SHOULDER HUMERAL SOCKET REAMER, SMALL PF
Common Device Name
REAMER
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key19102526
Report Number1644408-2024-00458
Device Sequence Number7268656
Product Code HTO
UDI-Device Identifier00888912171151
UDI-Public00888912171151
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number804-06-005
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer05/03/2024
Initial Report FDA Received Date04/13/2024
Supplement Report FDA Received Date06/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age72 YR
Patient SexFemale
Patient Weight113 KG
-
-