• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JIANGSU CAINA MEDICAL CO., LTD. MCKESSON; SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JIANGSU CAINA MEDICAL CO., LTD. MCKESSON; SYRINGE Back to Search Results
Catalog Number 16-S30C
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned samples and retained samples were inspected, all of them met the specification, the dhr of this lot was reviewed without any issuses.The root cause can't detected currrently, no action will be taken.This issue will be monitored.A supplemental report will be submitted if further information received.
 
Event or Problem Description
The customer report that barrel rupture.This defect was recognized when the drug was injected into the saline bag after it was drawn into the syringe.The sample on the feedback photo has a crack in the 5ml-20ml region.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON
Common Device Name
SYRINGE
Manufacturer (Section D)
JIANGSU CAINA MEDICAL CO., LTD.
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441 5
CH  214415
Manufacturer (Section G)
JIANGSU CAINA MEDICAL CO., LTD.
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441 5
CH   214415
Manufacturer Contact
camel zhou
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441-5
CH   214415
MDR Report Key19102877
Report Number3005670221-2024-00048
Device Sequence Number7428055
Product Code FMF
UDI-Device Identifier10612479206943
UDI-Public0110612479206943
Combination Product (Y/N)N
PMA/510(K) Number
K113091
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number16-S30C
Device Lot NumberCJCB03-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2023
Initial Report FDA Received Date04/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-