• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JIANGSU CAINA MEDICAL CO., LTD. MCKESSON; INSULIN SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JIANGSU CAINA MEDICAL CO., LTD. MCKESSON; INSULIN SYRINGE Back to Search Results
Catalog Number 102-SN1C31516P
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the insulin leak from the syringe, a big percentage in a case have this leakage.
 
Additional Manufacturer Narrative
The returned samples was requested but hasn't been received till now, the lot number of this event was received, the dhr of this lot was reviewed without any issuses.The retained samples of the lot was tested and the samples met the specification.The root cause can't detected currrently, no action will be taken currently, a supplemental report will be submitted if further information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON
Common Device Name
INSULIN SYRINGE
Manufacturer (Section D)
JIANGSU CAINA MEDICAL CO., LTD.
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441 5
CH  214415
Manufacturer (Section G)
JIANGSU CAINA MEDICAL CO., LTD.
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441 5
CH   214415
Manufacturer Contact
camel zhou
no 23, huanxi rd
zhutang town
jiangyin, jiangsu 21441-5
CH   214415
MDR Report Key19102893
Report Number3005670221-2024-00087
Device Sequence Number12619519
Product Code FMF
UDI-Device Identifier20612479187911
UDI-Public(01)20612479187911
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151949
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/31/2026
Device Catalogue Number102-SN1C31516P
Device Lot NumberCJCA04-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2021
Initial Report FDA Received Date04/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-