EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 3 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-35-3013 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10.Pending investigation.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a left total knee arthroplasty on (b)(6) 2011, and then experienced revision surgical procedure on (b)(6) 2022 approximately 10 years and 10 months after initial implant.Preoperative diagnosis: aseptic loosening, osteolysis, poly wear.No images were provided.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.H6: corrected health effect, medical device, component, and investigation clinical codes.
|
| |
|