EXACTECH, INC. OPTETRAK TOTAL KNEE SYSTEM, HI-FLEX ASYMMETRIC POSTERIOR-STABILIZED; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-35-1009 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
01/05/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a right total knee arthroplasty on (b)(6) 2015, with no reported revision.No other patient information / medical history reported.No radiographic images of the device are provided.The device will not be returned.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10.H10.Pending investigation.These devices are used for treatments, not diagnosis.There is no other information available.Concomitant products: 3668894 - 02-010-01-0310 - logic femoral ps cem right sz 1 2107850 - 02-012-45-1010 - lgc tibial fit tray cem sz 1f / 1t 2328176 - 200-03-26 - one peg patella 26mm.
|
| |
|
Additional Manufacturer Narrative
|
|
A2: information added: date of birth.A3b: information added: gender.D1: corrected the following: brand name.G1: corrected the following: reporting contact first name, reporting contact last name, reporting contact phone number.H6: corrected the following: health effect - clinical code, impact code, component code, type of investigation, investigation findings, investigation conclusions.The reason for the reported event cannot be confirmed from the information provided but may be due to prosthesis wear and/or inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs and relevant clinical information were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|