| Model Number |
153905-002 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Midmark received a complaint of a light head post of a ceiling mounted led dental light reported to have come loose causing the post, and presumably also the light, to fall.No injury to patient or staff was involved.Due to previous reporting of this or similar light falling events, midmark complied to report this instance.
|
| |
|
Additional Manufacturer Narrative
|
|
No return was issued for the device involved in this event.Due to the information provided at the time of this report, midmark corporation believes the root cause of this event is set up error.
|
| |
|