| Model Number |
CR2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer received a replacement lid and battery to resolve the reported issue.The device was not returned to stryker for evaluation.The root cause was unable to be determined.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that the magnet was missing from the lid of their device.This will result in an inability of the device to detect the lid being opened or closed.In this state, a delay in defibrillation may occur, as the users has to push the power button underneath the lid to turn the device on.There were no reports of patient use associated with the reported event.
|
| |
|
Search Alerts/Recalls
|