Model Number G407209
Medical Device Problem Codes Difficult to Insert (1316); Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 04/08/2024
Type of Reportable Event Serious Injury
Event or Problem Description
During a pulmonary vein isolation procedure, a small blood clot was noted on the needle.Transeptal puncture (with a stylet) was difficult and required multiple attempts to cross the septum.After one of the attempts, the needle was removed from the patient and it was noticed that then needle could not be flushed or drawn back.The physician asked for another needle and the case continued without any adverse events.The original needle was later checked outside the patient by reinserting the dilator and a small blood clot was pushed out of the needle and the needle could be flushed easily.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported blood clot remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key19105986
Report Number3008452825-2024-00200
Device Sequence Number18076741
Product Code DRC
UDI-Device Identifier05414734205184
UDI-Public(01)05414734205184(10)10012539(17)260930
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberG407209
Device Lot Number10012539
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2024
Supplement Date Received by Manufacturer04/25/2024
Initial Report FDA Received Date04/15/2024
Supplement Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight94 KG
-
-