| Catalog Number |
03183696122 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The c501 module serial number was (b)(6).
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of discrepant results for 1 patient sample tested for iron gen.2 (iron2) on a cobas 6000 c501 module.The initial result was 15.30 ug/dl.The sample was repeated 3 times with results of 25.90 ug/dl, 16.50 ug/dl, and 16.10 ug/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and qc were acceptable, therefore, a general reagent issue can be excluded.Based on the information provided, the investigation determined the event was consistent with a preanalytical handling issue.The investigation did not identify a product problem.
|
| |
|