| Model Number |
PM7100 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a "coronary artery bypass warm beating ," the sensor had erratic rso2 readings from below 15 to 50s or no readings.The monitor, preamplifier, and cables were swapped out, but the sensor continued to read erratically.The perfusionist placed the sensor on themselves and got normal somatic readings with no issues.The patient could not be monitored, and there was a delay in the procedure.It was reported that the procedure was not conducted because rso2 were not reliable.The delay in procedure did not cause harm to patient.
|
| |
|