• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRIA FIRM; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRIA FIRM; STENT, URETERAL Back to Search Results
Model Number M0061902330
Medical Device Problem Codes Difficult to Advance (2920); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable event of stent kinked, inside the patient.
 
Event or Problem Description
It was reported to boston scientific that a tria firm ureteral stent was used during a ureteroscopy procedure in the ureter performed on (b)(6) 2024.During insertion, when the stent was advanced through the guidewire, it was noted that they had difficulty advancing the stent.The stent was removed and found some folds in the coil part.The procedure was completed with another of different size of tria firm ureteral stent.There were no patient complications reported.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable event of stent kinked, inside the patient.Block h10: the returned tria ureteral stent was analyzed, and upon magnification and media inspection it was observed that the renal and bladder coil kinked.Additionally, the suture and positioner were not returned.The mandrel 0.039 was loaded into the device and no resistance was felt.No other problems with the device were noted.The reported event was confirmed.With all the available information, boston scientific concludes that the reported event was confirmed.It is possible to conclude that the kinks may have been generated due to excessive manipulation of the stent before being inserted over the guidewire.Once the device is bent, if the insertion continues, the kinks could generate difficulty in advancing over the guidewire.Therefore, all compiled information on this investigation determines that the most probable cause is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.
 
Event or Problem Description
It was reported to boston scientific that a tria firm ureteral stent was used during a ureteroscopy procedure in the ureter performed on (b)(6) 2024.During insertion, when the stent was advanced through the guidewire, it was noted that they had difficulty advancing the stent.The stent was removed and found some folds in the coil part.The procedure was completed with another of different size of tria firm ureteral stent.There were no patient complications reported.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIA FIRM
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19109916
Report Number2124215-2024-21133
Device Sequence Number15524746
Product Code FAD
UDI-Device Identifier08714729937913
UDI-Public08714729937913
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K190603
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/12/2026
Device Model NumberM0061902330
Device Catalogue Number1983-01
Device Lot Number0032415931
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/21/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date04/15/2024
Supplement Report FDA Received Date06/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-