MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Pain (1994)
Date of Event 03/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a knee revision approximately 16 days post implantation due to a periprosthetic fracture after a fall.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign - australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a right knee revision approximately 15 days post implantation due to pain and periprosthetic fracture from falling while walking to bathroom using a four-wheel walker and nursing assistance.All components exchanged with nexgen rotating hinge components without complications.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
No additional information to report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: proposed component code: mechanical (g04) - femur.The reported event was confirmed via provided medical records.Visual examination of provided photos identified explanted implants with foreign material covering the implants.Medical records were reviewed by a healthcare professional and identified the following: pre-revision, there was pain and x-ray confirmed supracondylar fracture above femur component.Revision op notes state that femur component and fracture fragment were excised and permanent components were placed with good rom and good stability.Radiographs were also provided and reviewed by a healthcare professional, which identified a comminuted and displaced periprosthetic fracture of the distal right femur, and severe osteopenia.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19111139
Report Number0001822565-2024-01240
Device Sequence Number12851903
Product Code MBH
UDI-Device Identifier00889024230767
UDI-Public(01)00889024230767(17)340202(10)66339306
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502806202
Device Lot Number66339306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2024
Supplement Date Received by Manufacturer04/15/2024
06/28/2024
Initial Report FDA Received Date04/15/2024
Supplement Report FDA Received Date05/06/2024
07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
-
-