EXACTECH, INC. LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 63yo male patient, initial left knee implanted on (b)(6) 2019, underwent a revision procedure on (b)(6) 2024, approximately 5 years post the initial procedure.The patient was revised due to femur loosening and poly wear.- everything was removed and exchanged to a competitor¿s product.There were no surgical delays or device breakages during the procedure.X-rays were provided.The patient was last known to be in stable condition following the event.No explanted devices are available for analysis.They were disposed of by the hospital.Device images were provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products: - logic cr femoral cem, left sz 2.5 (cat# 02-010-03-0225 / serial# (b)(6)).- lgc tibial fit tray cem sz 2.5f / 5t (cat# 02-012-45-2525 / serial# (b)(6)).- three peg patella 35mm (cat# 200-02-35 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was most likely the result of a combination of prosthesis wear secondary to contributions from implant alignment and/or instability of the knee as well as loosening due to mechanical loss of fixation at the cement/implant interface.However, this cannot be confirmed as the devices were not returned for evaluation and the loosening could not be confirmed on the pre-revision radiograph.
|
| |
|