| Model Number |
15 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A stryker service technician evaluated the customer's device and verified the reported issue.After replacing the main keypad assembly, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device would not power on.As a result, defibrillation therapy would not be available, if needed.There was no patient use associated with the reported event.
|
| |
|