Model Number CS1010C
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Pressure Sore/Ulcer (2326)
Date of Event 03/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer's wife claims that customer developed a stage 2 pressure sore due to cushion going flat but roho, inc.Has not seen medical records to confirm this.The manufacturing records were reviewed and indicated the cushion passed all inspections and verifications.The cushion was returned and a product evaluation was performed, which identified cushion creasing with probable cause due to improper product maintenance.The user has warning in the manual to discontinue use of a defective product that reads as follows: "check skin frequently, at least once a day.Redness, bruising, or darker areas (when compared to normal skin) may indicate superficial or deep tissue injury and should be addressed.If there is any discoloration to skin/soft tissue, stop use immediately.If the discoloration does not disappear within 30 minutes after disuse, immediately consult a healthcare professional.- check inflation frequently, at least once a day.- do not use a product that is underinflated or overinflated, because 1) the product benefits will be reduced or eliminated, resulting in an increased risk to skin and other soft tissue".If additional information is received, a follow-up report will be submitted.
 
Event or Problem Description
Customer's wife claims that customer developed a stage 2 pressure wound due to cushion going flat.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROHO® CONTOUR SELECT® CUSHION
Common Device Name
WHEELCHAIR CUSHION
Manufacturer (Section D)
ROHO, INC.
1501 s 74th st
belleville IL 62223
Manufacturer (Section G)
ROHO, INC.
1501 s 74th st
belleville IL 62223
Manufacturer Contact
cynthia jackson
1501 s 74th st
belleville, IL 62223
6182223542
MDR Report Key19113438
Report Number3008630266-2024-00002
Device Sequence Number7819661
Product Code KIC
UDI-Device Identifier00613732058517
UDI-Public0100613732058517
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberCS1010C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2024
Initial Report FDA Received Date04/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age37 YR
Patient SexMale
-
-