This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the biopsy inlet piece loose.Based on the result, we concluded that it was caused due to the excessive force applied on the biopsy inlet piece.In addition, our technician confirmed that the control body fluid damage, the forward body frame fluid damage, the biopsy inlet t-piece fluid damage, the forward body cover broken, the side body cover broken, the cfb (coherent fiber bundle) broken, the insertion flexible tube perforated, the insertion flexible tube buckled, the bending rubber stretched, the light source connector corroded, the insertion flexible tube crushed, and the operation channel (primary) resistance; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0620(control body)"" and/or the risk analysis results, it was evaluated to submit mdr.
|