MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Catalog Number AVSM07040
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reporter that during a stent graft placement procedure, the stent allegedly jumped it was being deployed and was allegedly ended up in the wrong spot.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.D4 (expiry date: 09/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation and no images were provided for review which leads to inconclusive result.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Holding and handling of the system throughout deployment was found sufficiently described.In particular the instructions for use state: 'do not touch the distal catheter assembly (i.E.The dark brown catheter segment) during covered stent deployment since this may interfere with covered stent deployment and may lead to misplacement' and 'slowly and carefully activate the covered stent release mechanism by rotating the large wheel.After deployment of approximately 15 mm, wait several seconds to allow the distal end of the covered stent to fully expand.Ensure the covered stent has wall apposition before completing deployment'.Regarding pta the instructions for use state: 'pre-dilate the stenosis with a pta balloon catheter'.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reporter that during a stent graft placement procedure in the severe stenosis of proximal portion of the right upper extremity arteriovenous graft via an antegrade approach, the stent allegedly jumped while it was being deployed and was allegedly ended up in the wrong spot.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation and no images were provided for review which leads to inconclusive result.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.Holding and handling of the system throughout deployment was found sufficiently described.In particular the instructions for use state: 'do not touch the distal catheter assembly (i.E.The dark brown catheter segment) during covered stent deployment since this may interfere with covered stent deployment and may lead to misplacement' and 'slowly and carefully activate the covered stent release mechanism by rotating the large wheel.After deployment of approximately 15 mm, wait several seconds to allow the distal end of the covered stent to fully expand.Ensure the covered stent has wall apposition before completing deployment'.Regarding pta the instructions for use state: 'pre-dilate the stenosis with a pta balloon catheter'.H10: b5, d4 (expiration date: 09/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reporter that during a stent graft placement procedure in the severe stenosis of proximal portion of the right upper extremity arteriovenous graft via an antegrade approach, the stent allegedly jumped while it was being deployed and was allegedly ended up in the wrong spot.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation and images were provided for review.Provided images demonstrate two overlappingly placed and expanded stents with a digitally drawn brace which may indicate the intended placement site.However, the image of a lying stent cannot confirmed stent jumping when the start position of the deployment is not known which leads to inconclusive result.It was only known that 6f introducer with 0.035" guidewire were used for access.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.Holding and handling of the system throughout deployment was found sufficiently described.In particular the instructions for use state: 'do not touch the distal catheter assembly (i.E.The dark brown catheter segment) during covered stent deployment since this may interfere with covered stent deployment and may lead to misplacement' and 'slowly and carefully activate the covered stent release mechanism by rotating the large wheel.After deployment of approximately 15 mm, wait several seconds to allow the distal end of the covered stent to fully expand.Ensure the covered stent has wall apposition before completing deployment'.Regarding pta the instructions for use state: 'pre-dilate the stenosis with a pta balloon catheter'.H10: d4 (expiration date: 09/2024), g3.H11: h6 (method).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reporter that during a stent graft placement procedure in the severe stenosis of proximal portion of the right upper extremity arteriovenous graft via an antegrade approach, the stent allegedly jumped while it was being deployed and was allegedly ended up in the wrong spot.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Common Device Name
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19113676
Report Number9681442-2024-00097
Device Sequence Number18067060
Product Code PFV
UDI-Device Identifier00801741106231
UDI-Public(01)00801741106231
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P170042
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAVSM07040
Device Lot NumberANGW1236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/29/2024
Supplement Date Received by Manufacturer07/10/2024
07/22/2024
07/31/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date07/20/2024
07/29/2024
08/07/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-