| Model Number |
ACU0T0 |
| Medical Device Problem Code |
Failure to Eject (4010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during an intraocular lens (iol) implant procedure, the lens was stuck in the injector.Additional information has been received and stated that, the procedure was completed with the another lens.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device and the lens were returned loose.A note with the complaint pr number was held in place by a rubber band onto the plunger spring of the device.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to mid-nozzle.No damage was observed to the device.The lens was returned adhered posterior surface up in dried solution on the exterior of the device.No damage was observed to the lens.Product history records were reviewed and documentation indicated the product met release criteria.A non-qualified viscoelastic was indicated.Root cause: the lens was outside the device when received.The root cause for the reported complaint may be related to a failure to follow the ifu.A non-qualified viscoelastic was indicated.The ifu instructs: during device preparation and implantation of the company iq iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: ¿ due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|