The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review implanting an expired device, not prepping the skin with antiseptic solution, not irrigating the skin with an antibiotic solution before closure, multiple tunneling attempts, using inappropriate tools, not using antibiotics pre-operatively, and the patient not attending their post-op visit have been ruled out as potential causes.Additionally, the patient did not touch or pick at their wound.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of the reported issue is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
|