| Model Number |
FR8A-RCV-A0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Electric Shock (2554)
|
| Date of Event |
03/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The unintended stimulation/new pain questionnaire was reviewed for potential causes of the reported issue.Based on this review changing the parameters of the transmitter has been ruled out as a potential cause.Per freedom scs instructions for use (05-20400), "a patient with the scs freedom-8a neurostimulator (electrode arrays with connected receiver) can be safety scanned in an mr system meeting the following conditions: static magnetic field of 1.5-tesla or 3-tesla." the cause of the reported issue is unknown.The stimulator is used to treat pain.The cause of the reported issue is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in unintended stimulation/new pain.Unintended stimulation/new pain issue rates remain acceptably low; thus, a capa is not required.Unintended stimulation/new pain issues rates will continue to be tracked and trended.
|
| |
|
Event or Problem Description
|
|
The patient reported shocking sensations on the right torso and down the thigh during a 3t mri.The patient has several total joint implants.Following the mri the patient still had slight pain and had not used the device since the event.The patient reached out inquiring if the device could still be used and if any damage was done.The patient was instructed to put the device back on and began receiving relief within 30 minutes.The patient is no longer experiencing slight pain and is satisfied with the pain relief they are currently receiving.
|
| |
|
Search Alerts/Recalls
|