• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURONIX LLC FREEDOM SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM SPINAL CORD STIMULATOR Back to Search Results
Model Number FR8A-RCV-A0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Electric Shock (2554)
Date of Event 03/15/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The unintended stimulation/new pain questionnaire was reviewed for potential causes of the reported issue.Based on this review changing the parameters of the transmitter has been ruled out as a potential cause.Per freedom scs instructions for use (05-20400), "a patient with the scs freedom-8a neurostimulator (electrode arrays with connected receiver) can be safety scanned in an mr system meeting the following conditions: static magnetic field of 1.5-tesla or 3-tesla." the cause of the reported issue is unknown.The stimulator is used to treat pain.The cause of the reported issue is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in unintended stimulation/new pain.Unintended stimulation/new pain issue rates remain acceptably low; thus, a capa is not required.Unintended stimulation/new pain issues rates will continue to be tracked and trended.
 
Event or Problem Description
The patient reported shocking sensations on the right torso and down the thigh during a 3t mri.The patient has several total joint implants.Following the mri the patient still had slight pain and had not used the device since the event.The patient reached out inquiring if the device could still be used and if any damage was done.The patient was instructed to put the device back on and began receiving relief within 30 minutes.The patient is no longer experiencing slight pain and is satisfied with the pain relief they are currently receiving.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM SPINAL CORD STIMULATOR
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer Contact
tena jimmerson
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key19114531
Report Number3010676138-2024-00060
Device Sequence Number7820191
Product Code GZB
UDI-Device Identifier00818225020693
UDI-Public(01)00818225020693(17)211201
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K182720
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/01/2021
Device Model NumberFR8A-RCV-A0
Device Lot NumberSWO191218
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2024
Initial Report FDA Received Date04/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-