SHIRAKAWA OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during reprocessing, the uretero-reno videoscope tested positive for 2 colony forming units (cfus) of aeromonas hydrophila.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Additional Manufacturer Narrative
New information added to the following fields: d8, d9.This report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the lms final investigation.Despite good faith attempts the user cleaning disinfection and sterilization (cds) processes were not shared.Olympus/ the user facility provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: <1cfu.Bacterial identification: n/a.The device was evaluated where the following abnormalities were found that could have led to the reported event; the device was repaired by a third party including the bending section cover adhesive, the insertion tube, the insertion tube protector, the scope connector and the switch button.As the device was repaired by a third party with non-olympus materials; the defined biocompatible characteristics are no longer guaranteed, for example; it was unknown if an improper adhesives was used or improper materials were used.Additionally, the defined mechanical characteristics are no longer guaranteed; for example the melting of improper insulation material instead of the ceramic manufactured by olympus.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." as a result of confirming the contents of the instruction manual, reprocess method is described on the items below; chapter 6 compatible reprocessing methods and chemical agents chapter 7 cleaning, disinfection, and sterilization procedures olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Common Device Name
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19115092
Report Number3002808148-2024-03525
Device Sequence Number15229603
Product Code FGB
UDI-Device Identifier04953170340321
UDI-Public04953170340321
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberURF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2024
Supplement Date Received by Manufacturer11/28/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-